• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. DESIGN OPTIONS; EPIDURAL ANESTHESIA KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. DESIGN OPTIONS; EPIDURAL ANESTHESIA KIT Back to Search Results
Model Number 552153
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: "the catheter knotted while placing cse.".
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).  no sample was returned for further evaluation.Retain samples were pulled and tested.All testing were within specification.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.  we will maintain this report for further references and continue to monitor other reports for similar occurrences.  if any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESIGN OPTIONS
Common Device Name
EPIDURAL ANESTHESIA KIT
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd
allentown 18109
MDR Report Key10810486
Report Number2523676-2020-00314
Device Sequence Number4011845
Product Code OGE
UDI-Device Identifier04046955214426
UDI-Public04046955214426
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2021
Device Model Number552153
Device Catalogue Number552153
Device Lot Number0061734344
Initial Date Received by Manufacturer 10/26/2020
Supplement Date Received by Manufacturer10/26/2020
Initial Report FDA Received Date11/09/2020
Supplement Report FDA Received Date12/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-