| Model Number |
EM800N |
| Medical Device Problem Codes |
Overheating of Device (1437); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report confirmed.The device was tested and the temperature was measured to be 150.3°f.The rate of temperature rise was above threshold at 6.8°f/sec.The likely cause was identified as worn bearing.It was also noted that the motor seized, the shaft was broken and the laser markings were illegible/faded.This type of failure is associated with (b)(4).The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with no reported malfunction.It was reported that there was no patient, or staff impact.Upon analysis, it was noted that the device overheated above threshold.On follow up, it was reported that no further information can be provided regarding the event.
|
| |
|