SMITH & NEPHEW, INC. CAMERA HEAD LENS INT SYS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72203967
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/15/2020
Event Type  malfunction  
Event Description
It was reported that during a shoulder procedure, the camera head had no image.Backup device was available to complete the procedure.No delay and no patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of no live video output was confirmed.Cause of video malfunction is a defective electronic component.The complaint investigation has concluded that the root cause of blank image is a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.Reference: case (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAMERA HEAD LENS INT SYS
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10813383
MDR Text Key215501097
Report Number1643264-2020-01737
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00885556629833
UDI-Public00885556629833
Combination Product (y/n)N
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72203967
Device Catalogue Number72203967
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Date Manufacturer Received02/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-