| Model Number |
NL473 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reference code no165z.Device name as univation xf tibia cemented t4.Lm.Serial number n/a.Batch number 52231095.Udi device identifier (b)(4).Udi production identifier (b)(4).Basic udi-di n/a.Unit of use udi-di (b)(4).Manufacturing date 2016-04-20.Ref.Code device name batch nr.No188z as univation xf femur cemented f4 lm 52235667.Nl473 univation f meniscal comp.T4 rm/lm 7mm 52232908.Nx044 patella 3-pegs p4 52316618.Np583r threaded pin headless 3.2mm x 63mm unknown.Investigation.No product at hand.Therefore an investigation at the product is not possible.Pictorial documentation.There are no pictures available.Batch history review.The device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.There are 2 similar complaints against the same lot number 52232908.(all registered as involved component).Explanation and rationale.In the light of the little information received and due the circumstance that we did not received the complained devices it is not possible to determine a root cause for the mentioned failure.Corrective action.For this topic (vega loosening) a product safety case (psc) was initiated.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an issue with knee components.As a result of having the product implanted the patient has experienced unknown claims, but it is known that it resulted in a revision surgery.The primary surgery occurred on (b)(6) 2016 and the revision surgery occured on (b)(6) 2017.Cement used: zimmer palacos cement.All available information has been provided at this time, if additional information becomes available the complaint will be updated accordingly.The adverse event / malfunction is filed under reference (b)(4).Involved components: no165z (tibial cemented t4 lm).No188z (femur cemented f4 lm).Nl473 (pe insert t4 rm/lm 7mm).Nx044 (universal patella p4).Np583r (threaded pin, headless 3.2mmx63mm).Associated medwatches: 2916714-2020-00570.2916714-2020-00571.2916714-2020-00573.2916714-2020-00574.
|
| |
|
Search Alerts/Recalls
|