| Model Number |
BXA087902A |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional device information: bxa087902a serial number: (b)(4) udi: (b)(4) as it is not known which device became separated from the other, both devices are being investigated.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020 a patient underwent an endovascular procedure in the (b)(6) study.Gore® viabahn® vbx balloon expandable endoprostheses were utilized in the branch vessels.On (b)(6) 2020 imaging revealed that the two vbx devices which were deployed in the superior mesenteric artery had become separated.On (b)(6) 2020 a reintervention to treat the type iii endoleak was performed.An additional vbx was deployed to bridge the separation and the endoleak was considered resolved.The patient tolerated the procedure.On (b)(6) 2020 the patient was discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.H.6.Results code 1: 213: a review of the manufacturing records for the device verified the lot met pre-release specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Results code 1: 213 a review of the manufacturing records for both devices verified the lots met all pre-release specifications.
|
| |
|