• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO NC EUPHORA RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO NC EUPHORA RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number NCEUP3020X
Medical Device Problem Codes Burst Container or Vessel (1074); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure an attempt was made to use one nc euphora rx ptca balloon catheter to treat a moderately tortuous, severely calcified lesion exhibiting 90% stenosis located in the proximal/mid right coronary artery (rca).The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.It was reported that the balloon burst at 20atm, post orbital atherectomy.The patient was reported to be alive with no injuries.
 
Additional Manufacturer Narrative
Additional information: the lesion was pre-dilated by the device.The device was not moved or repositioned while inflated.It was stated that two inflation were performed prior to the burst occurring.The balloon burst occurred on the 3rd inflation.The patient was reported to be alive with no injuries.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
1 still image received image provided is of the distal section of a poba device with what looks to be a radial burst.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis summary: the device returned with kinks visible on the hypotube and distal shaft/support wire.A complete radial burst was evident on the balloon, causing a detachment of the balloon material.It appears to be a clean detachment on the proximal side of the burst, while the material was distorted on the distal side of the burst.Blood visible in the balloon.The inner lumen patency could not be verified with a 0.015 inch mandrel due to kinking on the distal shaft.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EUPHORA RX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key10815261
Report Number9612164-2020-04332
Device Sequence Number1926955
Product Code LOX
UDI-Device Identifier00643169395848
UDI-Public00643169395848
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K141090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2022
Device Model NumberNCEUP3020X
Device Catalogue NumberNCEUP3020X
Device Lot Number220625248
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Initial Date Received by Manufacturer 11/09/2020
Supplement Date Received by Manufacturer11/10/2020
01/13/2021
01/21/2021
Initial Report FDA Received Date11/10/2020
Supplement Report FDA Received Date12/02/2020
01/20/2021
02/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age84 YR
Patient Weight82
-
-