• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LYNX BLUE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LYNX BLUE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR Back to Search Results
Model Number M0068503010
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a lynx blue system device was used during a sling procedure performed on (b)(6) 2020.According to the complainant, part of the plastic sleeve detached during the procedure.The detached piece did not fall off into the patient.The procedure was completed with another lynx blue system device.There were no patient complications reported as a result of the event.
 
Additional Manufacturer Narrative
Block h6: device code 2907 captures the reportable event of sleeve detachment.Block h10: the mesh sleeve was returned without the mesh or the delivery devices.A visual examination of the returned sleeve revealed that it was cut in two.The association loops were present on both ends of the sleeve and there was no evidence that there were pieces missing.Returned product analysis was unable to identify any issue with the returned components.Therefore, the complaint could not be confirmed.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.Based on the investigation results, the most probable cause chosen for this complaint is no problem detected as there was no evidence found to suggest an issue with the device.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a lynx blue system device was used during a sling procedure performed on (b)(6), 2020.According to the complainant, part of the plastic sleeve detached during the procedure.The detached piece did not fall off into the patient.The procedure was completed with another lynx blue system device.There were no patient complications reported as a result of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LYNX BLUE SYSTEM
Common Device Name
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10815519
Report Number3005099803-2020-05040
Device Sequence Number19361388
Product Code OTN
UDI-Device Identifier08714729961932
UDI-Public08714729961932
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K020110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2023
Device Model NumberM0068503010
Device Catalogue Number73193
Device Lot Number0025376844
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2020
Initial Date Received by Manufacturer 10/12/2020
Supplement Date Received by Manufacturer12/08/2020
Initial Report FDA Received Date11/10/2020
Supplement Report FDA Received Date01/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-