Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for all available lot numbers; the devices were found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
|
It was reported to aesculap inc.That an vega knee implant system, comprised of the following components, vega system ps femur f5n r (part #nx031z) (ref.Associated mdr id 2916714-2020-00521), vega system patella 3-peg size p3 (part # nx043) (ref.Associated mdr id 2916714-2020-00522), vega system ps tibia plateau t3, cemented implant (part # nx055z), ps+ gliding surface implant 10mm t3/t3+ (part # nx130) (ref.Associated mdr id 2916714-2020-00524), and a vega system peek plug t1-t5 (part # nn260p) (ref.Associated mdr id 2916714-2020-00525), was implanted during a primary surgery performed on (b)(6) 2013.According to the complainant, following the primary surgery, the patient experienced pain and swelling in right knee, which restricted routine daily activities.A subsequent x-ray showed possible loosening of the tibial component and a radiolucent line around the tibial component.Bone scans showed an increased uptake in the tibial component.Intraoperative findings during replacement surgery were a well-fixed patellar component and losing of the tibial and femoral components.The cement used during the primary surgery was identified as zimmer palacos r radiopaque bone cement.The complaint device is not available to be returned to the manufacturer for evaluation.Additional information has been requested, but has not been made available.The adverse event / malfunction is filed under aag reference (b)(4).Involved components: nx031z - as vega ps femoral comp.Cemented f5n r - lot # 51931511, nx043 - patella 3-pegs p3 - lot # 51955964, nx055z - as vega ps tibial plateau cemented t3 - lot # 51938561, nx130 - vega ps gliding surface t3/3+ 10mm - lot # unknown, nn260p - plug f/tibial plateau - lot # 51974495.
|