| Model Number |
2842 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Chest Pain (1776)
|
| Date of Event |
10/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that guidewire fracture occurred and the patient experienced chest pain.The 95% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A 190cm, straight-tip samurai guidewire was advanced to cross the lesion.However, during the procedure, the guidewire fractured.It took a very long time to cross the lesion, once the wire crossed the lesion the patient started having chest pain so the physician opted to continue with using the same guidewire.The wire was removed intact and no patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for analysis.Examination of the returned product showed a large curve within approximately 15 mm from the distal tip and pitch elongation of 3 wraps of the coil.
|
| |
|
Event or Problem Description
|
|
It was reported that guidewire fracture occurred and the patient experienced chest pain.The 95% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A 190cm, straight-tip samurai guidewire was advanced to cross the lesion.However, during the procedure, the guidewire fractured.It took a very long time to cross the lesion, once the wire crossed the lesion the patient started having chest pain so the physician opted to continue with using the same guidewire.The wire was removed intact and no patient injury was reported.
|
| |
|