• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1737
Device Problems Failure to Fire (2610); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems No Code Available (3191); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 09/18/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a procedure, the device did not work when activating the coagulation button.The generator was replaced but it still did not work.The surgery started by video with the device indicated for this surgery, however with the non-functioning of the device, it was not feasible to proceed with the video surgery, so the surgery was converted to the conventional "open" form.
 
Manufacturer Narrative
Additional information: d8, d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was an adverse event without an identified device or use problem.A potentially related device issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.It was also reported that the device did not activate when keyed.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key10816696
MDR Text Key215509230
Report Number1717344-2020-01385
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10884521181571
UDI-Public10884521181571
Combination Product (y/n)N
PMA/PMN Number
K133338
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/16/2022
Device Model NumberLF1737
Device Catalogue NumberLF1737
Device Lot Number73110156X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Date Manufacturer Received02/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
Patient Weight60
-
-