| Model Number |
M00556581 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(initial reporter address): (b)(6).(b)(4).A visual examination of the device revealed the ptfe peeled in the middle section of the jagwire,therefore, the complaint was confirmed.In addition, the guidewire was bent at the distal section.It is likely that the failure modes observed have been caused due to interaction with the sphincterotome or when the jagwire was being inserted into a non boston scientific duodenoscope device.If extra force was applied during usage of the jagwire or if the guidewire was exposed to the voltage applied to sphincterotomy device during procedure, this could have lead to the ptfe peeling.Based on the information available and the analysis performed, the investigation conclusion code for the complaint will be documented as adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed, and from the information available this device was used per the instructions for use (ifu).
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that jagwire guidewire was used in the biliary system (papilla) during an endoscopic retrograde cholangiopancreatography (ercp) procedure on (b)(6) 2020.According to the complainant, during the procedure, when papilla was cut with a sphincterotome and guidewire was inserted in the non boston scientific duodenoscope, the ptfe peeled.This happened with seven jagwires.The procedure was cancelled based upon no additional stock available.There were no reported patient complications as a result of this event.
|
| |
|
Search Alerts/Recalls
|