• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 14.0MM THREADED SPINDLE 480MM; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 14.0MM THREADED SPINDLE 480MM; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 394.40
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, patient underwent an open reduction internal fixation (orif) of right tibial plateau.During the procedure the nut on the femoral distractor got stuck in the middle of the bar.It got cross threaded either when it was being put into position or when it was being taken off.During the procedure they were not able to disconnect it; even using pliers.After the procedure, the bolt was able to be removed with some struggle.It was noted when the device had been assembled before the procedure, it was a little stiff, but it went together nicely.The procedure was successfully completed with no delay reported.This report is for a threaded spindle.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part 394.40, lot 34p1822: manufacturing site: bettlach.Release to warehouse date: june 10, 2020.A manufacturing record evaluation was performed for the part: 2801, lot: 34p1822, and no non-conformances were identified.H3, h6: a product investigation was completed: the device was received with the threads stripped.No other issues were identified with the returned components of the device.Dimensional inspection could not be conducted due to post manufacturing damage.The drawing, reflecting the manufactured and current revision, were reviewed.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.A review of the manufacturing record evaluations was performed for the finished device lot number, and no non-conformance were identified.The complaint condition is confirmed as the threads were stripped.During this investigation no product design or manufacturing issues were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
14.0MM THREADED SPINDLE 480MM
Common Device Name
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10817915
Report Number2939274-2020-05069
Device Sequence Number1114855
Product Code HST
UDI-Device Identifier10886982199683
UDI-Public(01)10886982199683
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number394.40
Device Catalogue Number394.40
Device Lot Number34P1822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Initial Date Received by Manufacturer 10/15/2020
Supplement Date Received by Manufacturer12/03/2020
Initial Report FDA Received Date11/10/2020
Supplement Report FDA Received Date12/08/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SPINDLE NUT; SPINDLE NUT
-
-