| Model Number |
401763 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
10/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro us drug-eluting stent system was selected for treatment.The proximal lad was uneventful and successfully stented with the orsiro us stent.There was an abrupt closure of a 2mm diagonal vessel with stent implantation, which resulted in some chest discomfort and minimal st elevation in the lateral leads.This was expected to be temporary with expected return of flow.After procedure patient experienced chest pain.A stent thrombosis was detected approx 2 hours after implantation.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the complaint instrument nor the angiographic material was available for analysis.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations no manufacturing related root cause was determined.However, it is noted that the patient vomited within the first two hours of recovery, although no pills were noted in the emesis, according to nursing.This may have been weakened the antiplatelet therapy.
|
| |
|