Model Number FA-55160-1030
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was received as medwatch (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the marksman catheter tip sheared during procedure.The issue was noticed after angiogram and thrombectomy when the catheter was removed.The broken portion was removed intact with no patient complications.
 
Event or Problem Description
Additional information received indicated that the devices were prepared per the ifu, the patient's vessel tortuosity was moderate, and the patient did not experience any symptoms as a result of the event.No further intervention was taken as the marker pieces unraveled but were still intact in the catheter.The cause of the event remains unknown.Ancillary devices include 6f shuttle jet 7, synchro guidewire.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
B5.Product analysis findings: the marksman total length was measured to be ~169.0cm and the useable length was measured to be ~161.5cm which is within specification (specification: 160cm ± 3cm).No damages or irregularities were found with the marksman hub.The marksman catheter body was fond kinked at ~23.0cm, 18.0cm, and 7.5cm from the distal end.The distal ~5.5cm of the marksman catheter body was found flattened.The marksman distal marker was found dislodged from the catheter tip.The dislodged marksman marker band was not returned.The separated outer tubing material exhibited with plastic deformation (jagged edges and stretching).There does not appear to be any separation of the inner liner.The inner wire was found stretched out from between the inner and outer tubing.No other anomalies were observed.Based on the device analysis and reported information, the customer¿s report of ¿catheter separation/break¿ was confirmed.The separated outer tubing material exhibited with plastic deformation which indicates the marker band dis lodged as the tensile strength of the tubing material was exceeded.The investigation determined that this event was similar to an event that had already been investigated, and another investigation is not necessary.Based on the formal investigation conducted possible causes for the event could be attributed to patient¿s highly tortuous anatomy and/or other contributing factors such as kink/damage in guide catheter / balloon guide catheter /guidewire/stents which may have contributed to resistance during delivery and/or retrieval, subsequently causing the marksman catheter to become damaged.For the marker band to dislodge from the catheter the distal shaft needs to be damaged and/or separated from the catheter, which is unlikely to occur without experiencing excessive resistance.In addition, hard clots /calcifications in the navigation tract can snag the catheter, this may be detected under fluoroscopic use of the product and as instructed in the ifu, without observing the resultant tip response, the catheter shouldn¿t be attempted to move further.As the marksman was found flattened it is likely that the guide catheter was kinked/damaged causing the marksman to flatten if pulled through, likely against resistance, subsequently causing the marker to dislodge.H6.Coding updated based on analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARKSMAN
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10819740
Report Number2029214-2020-01117
Device Sequence Number11291028
Product Code KRA
UDI-Device Identifier00847536026278
UDI-Public00847536026278
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2023
Device Model NumberFA-55160-1030
Device Catalogue NumberFA-55160-1030
Device Lot Number220523545
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2021
Initial Date Received by Manufacturer 10/28/2020
Supplement Date Received by Manufacturer01/05/2021
04/14/2021
Initial Report FDA Received Date11/10/2020
Supplement Report FDA Received Date01/14/2021
04/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient Weight63
-
-