BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110137
Device Problems Defective Device (2588); Mechanical Jam (2983)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2020
Event Type  malfunction  
Event Description
It was reported that when spd was cleaning the hand piece, they stated the gears would not turn.They lubed them, and manually tried to get them to turn without any luck.Not during surgery.No delay.No patient harmed.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The navio handpiece, part number pfsr110137, serial (b)(6) and intended to be used for treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was confirmed.The gear would not turn by hand.A handpiece test was then performed.The gears were able to turn by hand after the test.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most probable cause of the reported problem is early stage motor failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10819776
MDR Text Key216204266
Report Number3010266064-2020-01987
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
PMA/PMN Number
K170360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/21/2020
Initial Date FDA Received11/10/2020
Supplement Dates Manufacturer Received10/11/2021
Supplement Dates FDA Received10/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NPFS02000
-
-