• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS GO; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS GO; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66800603
Medical Device Problem Code Failure to Charge (1085)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/06/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the renasys is not charging.No injury to patient reported due this event.Unknown how the treatment was finished.
 
Additional Manufacturer Narrative
H10.The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key10819804
Report Number8043484-2020-03818
Device Sequence Number1114872
Product Code OMP
UDI-Device Identifier00040565125299
UDI-Public00040565125299
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K152163
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66800603
Device Catalogue Number66801496
Initial Date Received by Manufacturer 10/16/2020
Supplement Date Received by Manufacturer01/11/2021
Initial Report FDA Received Date11/10/2020
Supplement Report FDA Received Date01/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-