| Model Number |
DBP-125SOLID145 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Separation (1562)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/19/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The result of the event was amputation; however, the patient was already scheduled for amputation prior to use of oa.Oa was considered the last effort prior to going forward with amputation.(b)(4).
|
| |
|
Event or Problem Description
|
|
A diamondback peripheral orbital atherectomy device (oad) was used for treatment of a lesion in the anterior tibial artery (at).The patient was scheduled for amputation, but oa treatment was selected as a final effort to avoid amputation.The physician was aware that atherectomy could be challenging.The oad became stuck in the vessel during use.The wire was removed from the oad.The driveshaft of the oad was cut near the strain relief.A non-csi wire was advanced into the at in an attempt to free the crown.The driveshaft of the oad fractured, and the fragment remained in vivo.The patient was stable and underwent the previously planned amputation.
|
| |
|