| Model Number |
SUPERNOVA NASAL PAP VENTILATION SYSTEM (MEDIUM) |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical was able to confirmed the reported issue, and traced that the root cause was related to the manufacturing process supplier manufacturing process.As capa, all the personnel involved were retrained on spm snl-20, and a quality notification was created, and sent to our external supplier.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that during clinical settings, the ssm-20 supernova nasal pap ventilation system (medium) was not inflating, and holes were detected.As a resolution, the bag was replaced, and there was no patient harm reported.
|
| |
|