| Model Number |
UE00N59 |
| Medical Device Problem Codes |
No Display/Image (1183); Display or Visual Feedback Problem (1184); Output Problem (3005)
|
| Health Effect - Clinical Code |
Brain Injury (2219)
|
| Date of Event |
10/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation concluded that the network connection on the radiologist's computer is substandard for the use of viewing diagnostic images in a timely manner.Due to the long content download times, it was recommended that the radiologist wait until that series is mostly downloaded (identified with a green bar) before scrolling through the series with a mouse wheel.Manufacturer identified that when user pressess the up/down/left/right arrow keys to advance frames, the issue is not present.Supplemental report will be filed upon receipt of new and/or relevant information.
|
| |
|
Event or Problem Description
|
|
Customer reported frame skipping issues when playing cine series.The radiologist recently missed a brain bleed on a patient which caused a negative outcome, but did not result in death.The brain bleed apparently showed on only one image and it was the image that was skipped.Information reported indicates a reportable malfunction data display issue (non-image quality), which may cause or contribute to a serious injury requiring medical intervention; therefore, a reportable malfunction will be filed.
|
| |
|