H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one unopened monopty device was returned for evaluation.During visual evaluation, the sample was clean and the stylet and cannula were in their initial positions.Upon visual and microscopic evaluation, foreign material was noted within the packaging.No other visual anomalies noted on the device.Functional testing was not performed due to the condition of the device.Therefore, the investigation is confirmed for the alleged foreign material, as foreign material was noted with in the package.The definitive root cause could not be determined based upon the available information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 12/2020).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|