• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN LOCATOR ABUTMENT 4.1MM(D) X 4MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN LOCATOR ABUTMENT 4.1MM(D) X 4MM(H); DENTAL ABUTMENT Back to Search Results
Model Number ILOA004
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Information (3190)
Date of Event 08/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).
 
Event or Problem Description
It was reported that the abutment screw fractured and was removed.The doctor reported a delay of 30 minutes , new appointment would be required to place a new abutment.Gingiva was exposed.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).The following sections have been updated: h3: device evaluated by manufacturer: change ¿no' to 'yes.' the complaint report was received, reviewed, and evaluated by the manufacturer.Applicable manufacturer's internal lot history records were reviewed to evaluate whether any internal deviations, non-conformances, or problems occurred during the manufacture of the lot, which could cause or contribute to the reported complaint condition.Condition of the returned complaint related product was evaluated to confirm the reported product complaint.As part of each complaint investigation, the manufacturer performs an analysis to determine the root-cause of the reported problem.Depending upon the information provided, a true root-cause may or may not be determined.In these cases, the most-likely cause of the reported issue would be determined.Other known issues that are inherent to the business operations, typically ones that are related to shipping, handling, customer service, or other, are internally tracked and continually trended for unusual increases.As a result of the above investigation, the manufacturer has concluded the following as the root-cause or most likely root-cause of the reported complaint condition: no manufacturing defect was found most likely (see below): hazard experience thread fracture during function.Most likely root cause of the failure mode / hazard typically is the result from insufficient tightening torque at placement, overloading of the abutment in function, or not retightening the abutment at prescribed intervals.Known issue is being tracked and trended with no further actions taken at this time.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN LOCATOR ABUTMENT 4.1MM(D) X 4MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10822840
Report Number0001038806-2020-01792
Device Sequence Number3569514
Product Code DZE
UDI-Device Identifier00844868006091
UDI-Public(01)00844868006091(17)240212(10)1225492(241)ILOA004
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K012911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/12/2024
Device Model NumberILOA004
Device Catalogue NumberILOA004
Device Lot Number1225492
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/19/2020
Supplement Date Received by Manufacturer01/04/2021
Initial Report FDA Received Date11/11/2020
Supplement Report FDA Received Date01/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
-
-