• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 201101
Medical Device Problem Codes Mechanical Problem (1384); Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed a bend 5 cm from the distal tip.The handle, steering knob, and electrodes appeared to be intact.The device did not meet the curve and planarity specification.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: do not introduce the tip folded into the guiding sheath.Do not exert excessive pressure during placement of catheter if unknown resistance is encountered.Do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.Avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported the catheter does not make proper curve when turning.There was no medical intervention.The complainant is not aware of any patient injury, or extended procedure time.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS PHOENIX
10232 s. 51st st.
phoenix AZ 85044
MDR Report Key10823096
Report Number0002090040-2020-00033
Device Sequence Number3760053
Product Code NLH
UDI-Device Identifier00885825003845
UDI-Public00885825003845
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K030187
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2022
Device Model Number201101
Device Catalogue Number201101RR
Device Lot Number4182897
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2020
Initial Report FDA Received Date11/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-