| Model Number |
DTBA1Q1 |
| Medical Device Problem Codes |
High impedance (1291); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id 6932-65 lead, implanted: (b)(6) 1998; 429888 lead, implanted: (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the day after the patient¿s generator changout procedure the right ventricular (rv) lead exhibited high and undefined pacing impedance.An x-ray confirmed that the rv pace/sense lead was not inserted past the pin block of the implantable cardioverter defibrillator (icd).Reprogramming was done and the physician requested frequent monitory as the patient will be scheduled for a revision in two months to decrease risk of infection from early intervention.
|
| |
|