SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 3-4 12MM RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74035234
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/26/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the journey ii bcs art insert trial size 3-4 12mm right has developed a slight crack.There was no case involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).Results of investigation: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured on the mating surface preventing the device from performing as designed.The device was manufactured in 2013 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS ART ISRT TRL SZ 3-4 12MM RT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10823413
Report Number1020279-2020-06371
Device Sequence Number3204499
Product Code JWH
UDI-Device Identifier00885556157589
UDI-Public00885556157589
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74035234
Device Catalogue Number74035234
Device Lot Number13JAP5607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Initial Date Received by Manufacturer 10/26/2020
Supplement Date Received by Manufacturer11/19/2020
Initial Report FDA Received Date11/11/2020
Supplement Report FDA Received Date11/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-