| Model Number |
RMU-1000 |
| Medical Device Problem Code |
Failure to Run on Battery (1466)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
10/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, the device associated with this event has not been returned, and the cause for the event is not known.Should additional information become available, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
An emergency medical transportation company reported that their chest compressor stopped compressions during a rescue attempt, and that both the service and warning lights were lit.It was also reported that the patient was not resuscitated, and no additional event or device information was provided.It is not known if the device was removed and if manual compression were performed; therefore, this mdr is being submitted out of an abundance of caution.Although requested, the device associated with this event has not been returned and the cause for the event is not known.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation and testing of the returned chest compressor identified extensive internal rust, thick residue, debris and viscus fluids as well as mechanical damage in the device consistent with the software evaluation which shows numerous motor errors such as: "driver faulted at power up" and "motor did not move in time".
|
| |
|