| Model Number |
801-37545M |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that after final locking, the rod was still able to slide in the head of the mesa deformity uniplanar screw.Surgery was successfully completed with another screw, and an unknown delay.No adverse consequences, or medical intervention were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection: the returned screw had loss of anodization and slight deformation around the inner and outer collet which did not impact the functionality of the device.Functional inspection: the complaint screw was functionally inspected using the returned denali rod as well as a sample rod.The device was able to final lock onto the rods, using a sample quick locker, without issue.The rods were not able to slide along the collets, as reported.Device and complaint history records were reviewed for this lot, and no relevant manufacturing issues or similar complaints were identified.It is possible that the screw was not in a completely locked state during the procedure, causing the rod to slip along the screw head.As certain maneuvers applied during the procedure could not be replicated during functional testing, the failure mode could not be confirmed and the root cause could not be determined conclusively.
|
| |
|
Event or Problem Description
|
|
It was reported that after final locking, the rod was still able to slide in the head of the mesa deformity uniplanar screw.Surgery was successfully completed with another screw and an unknown delay.No adverse consequences or medical intervention were reported.
|
| |
|
Search Alerts/Recalls
|