It was reported that a patient was unstable, a revision surgery was performed to remove and replace the lgn porous cr fem sz 7l and legion porous ha tibial base sz 6 lt with a stryker revision.No other information available.Patient outcome is unknown.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the patient underwent a knee revision due to instability.Per email communication, the requested surgical reports and x-rays are not available.Therefore, the patient impact beyond the revision cannot be determined.Due to the limited information provided, no further clinical assessment can be rendered at this time.Should additional clinically relevant documentation become available, the clinical/medical task may be re-opened.A review of complaint history did not reveal additional complaints for the listed device.Device specific batch information was not provided.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to fit/sizing issue, lifetime of device or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|