• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER; KRA CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER; KRA CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number N/A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); No Code Available (3191)
Date of Event 11/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter occupation: unknown.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a male patient required placement of a cantata duo 2.8 superselective microcatheter for a splenic embolization procedure.After the coils were placed, the operator noted that the catheter broke inside of the patient, about 10cm from the distal end.The operator retrieved the separated portion with a snare.The procedure was completed successfully.No other adverse effects were reported for this incident.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Additional information: b5, d11 this report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Preliminary examination of the device returned 30nov2020 showed a kink/twist at the distal end of the catheter, but no evidence of separation.A small piece of metal/coiling that was snared from the patient as well as the competitor's snare were returned, however the metal is not a portion of the catheter.
 
Additional Manufacturer Narrative
Correction: this report is being sent to indicate the complaint event is not reportable.Upon further investigation and discussion with the customer, this event is not reportable.It was determined that the only failure of the catheter was that it was twisted/bent.An embolization coil was snared from the patient due to it being mistaken as a portion of the catheter, however no issue with the catheter necessitated "medical or surgical intervention to preclude permanent impairment of a body structure or permanent damage to a body stricture." a review of risk documentation does not indicate that this event is likely to cause serious injury if it were to reoccur.As there are no recorded incidences of serious injury due to the complaint event, and it is not likely that serious injury would result if the event were to recur, per 21cfr part 803.50 the complaint event is considered not reportable.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER
Common Device Name
KRA CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10825513
Report Number1820334-2020-02075
Device Sequence Number3016109
Product Code KRA
UDI-Device Identifier10827002238050
UDI-Public(01)10827002238050(17)220204(10)10317553
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K101450
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/04/2022
Device Model NumberN/A
Device Catalogue NumberMCS-2.8-NT2-135-15-HP
Device Lot Number10317553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Initial Date Received by Manufacturer 11/09/2020
Supplement Date Received by Manufacturer11/30/2020
01/18/2021
Initial Report FDA Received Date11/11/2020
Supplement Report FDA Received Date12/16/2020
01/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COOK EMBOLIZATION COILS
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age23 YR
-
-