• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, TRIFURC, ST; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, TRIFURC, ST; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H93863
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that holes were observed in the sterile primary packaging of three (3) trifurcated fluid transfer tube sets.This was identified prior to use.There was on patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added to d10, h3, h4 and h6.H4: the lot was manufactured from october 23, 2018 - october 24, 2018.H10: three (3) actual samples were received for evaluation.Unaided visual inspection was performed which observed a small hole in the primary packaging of all samples.The reported condition was verified.The cause of the condition was not determined, however, the most likely cause was due to an impact (unknown) causing the small hole in all of the sample packages.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBE SET, STD VOL, TRIFURC, ST
Common Device Name
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10825643
Report Number1416980-2020-07036
Device Sequence Number3800104
Product Code LHI
UDI-Device Identifier00085412476698
UDI-Public(01)00085412476698
Combination Product (Y/N)Y
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K062909
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2021
Device Catalogue NumberH93863
Device Lot Number60154303
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Initial Date Received by Manufacturer 10/16/2020
Supplement Date Received by Manufacturer11/23/2020
Initial Report FDA Received Date11/11/2020
Supplement Report FDA Received Date12/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-