| Model Number |
W4-7-3-MVI-3 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Ruptured Aneurysm (4436)
|
| Date of Event |
10/12/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and not returned to the manufacturer for evaluation.Intraprocedural or post-procedural images were not provided for review; therefore, the reported event cannot be confirmed.The instructions for use identifies aneurysm rupture as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during deployment of the web, blood appeared to be coming from the aneurysm into the subarachnoid space.The web was detached in the aneurysm and the bleeding stopped.Post-procedure, the patient was reported to be "fine" and is currently doing well.There was no sequela.There was no reported device malfunction.
|
| |
|