(b)(4).The opt844 interface is used to deliver humidified oxygen to patients.The opt844 consists of a lightweight delivery tube which is connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's nares.Method: the complaint opt844 nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photograph provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Result: visual inspection of the provided photograph revealed the nasal prong of the complaint cannula was broken at the headgear connection point.Furthermore, the manifold and tube was found assembled to the incorrect side.Conclusion: the damage observed is most likely due to overtightening of the head strap.The opt846 interface may have been disassembled from the manifold and tube and incorrectly reassembled by the caregiver or user.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is disposed of.The subject cannula would have met the specification at the time of production.The setup instructions in the user instructions which accompany the opt844 nasal cannula include the following steps: - "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." - "cannula can become unattached if not used with the head strap clip." - "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." the user instructions also contain the following warnings/cautions: - "do not crush or stretch tube, to prevent loss of therapy." - "failure to use the set-up described above can compromise performance and affect patient safety.".
|