• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 75445540
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a health care provider via a manufacturer representative regarding a patient with l4-s1 fusion for spinal therapy.It was reported that the screw broke.Implanted and replaced with a new screw.The device was replaced and there was no fragment left inside the patient.The patient had back pain as symptoms.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3:part # 75445540, lot # h5231866.Visual and optical examination did not identify any cracks, fractures or breakage.Multiple driver interface features with deformation, consistent with the seating of the rod and implant and removal of the bone screw.After visual and optical inspection, the implant appears to be capable of performing its intended function.No fault found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key10827416
Report Number1030489-2020-01637
Device Sequence Number3645004
Product Code KWP
UDI-Device Identifier00885074134017
UDI-Public00885074134017
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K042025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75445540
Device Catalogue Number75445540
Device Lot NumberH5231866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2020
Supplement Date Received by Manufacturer01/14/2021
Initial Report FDA Received Date11/12/2020
Supplement Report FDA Received Date01/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
Patient Weight64
-
-