Model Number E2608-6
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Burn(s) (1757)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, after an adenoid reduction procedure, patient sustained a partial thickness skin burn on the lip during the procedure.An incident which happened for one of pediatric patient few weeks ago due to potentially damaged monopolar suction coagulator.The surgeon confirmed that there was no noticeable burn in the immediate post op but it gradually developed which made the surgeon to suspect that there was a fault in the insulation.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, after an adenoid reduction procedure, patient sustained a partial thickness skin burn on the lip during the procedure.An incident which happened for one of pediatric patient few weeks ago due to potentially damaged monopolar suction coagulator.The surgeon confirmed that there was no noticeable burn in the immediate post op but it gradually developed which made the surgeon to suspect that there was a fault in the insulation.The surgeon confirmed that he had to refer the patient to a plastic surgeon and is under treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key10827764
Report Number3006451981-2020-00179
Device Sequence Number3525295
Product Code GEI
UDI-Device Identifier20884521039015
UDI-Public20884521039015
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K791752
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/07/2025
Device Model NumberE2608-6
Device Catalogue NumberE2608-6
Device Lot NumberS0AE003X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/29/2020
Supplement Date Received by Manufacturer11/20/2020
Initial Report FDA Received Date11/12/2020
Supplement Report FDA Received Date12/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-