| Model Number |
MI1200 SYNCHRONY |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is planned in (b)(6) 2021.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.No accident or trauma has been reported but per parents the user is wobbly on her feet.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.No accident or trauma has been reported but per parents the user is wobbly on her feet.The user has been re-implanted.
|
| |
|