MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1200 SYNCHRONY
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user's hearing performance with the device is affected.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is planned in (b)(6) 2021.
 
Event or Problem Description
The user's hearing performance with the device is affected.No accident or trauma has been reported but per parents the user is wobbly on her feet.
 
Additional Manufacturer Narrative
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Event or Problem Description
The user's hearing performance with the device is affected.No accident or trauma has been reported but per parents the user is wobbly on her feet.The user has been re-implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key10827998
Report Number9710014-2020-00656
Device Sequence Number16900441
Product Code MCM
UDI-Device Identifier09008737094089
UDI-Public(01)09008737094089(11)160929(17)190929
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMI1200 SYNCHRONY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/10/2020
Supplement Date Received by Manufacturer11/10/2020
11/10/2020
Initial Report FDA Received Date11/12/2020
Supplement Report FDA Received Date02/05/2021
08/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
-
-