• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7210812S
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/01/1901
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the vulcan generator displayed a power cal failure.No case was involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Additional Manufacturer Narrative
H10, h3, h6: the reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of power cal failure could not be reproduced.Product passed functional testing with no faults or errors.Power cal failure did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL VULCAN GENERATOR CE MARK
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10828039
Report Number1643264-2020-01798
Device Sequence Number19895206
Product Code GEI
UDI-Device Identifier03596010542182
UDI-Public03596010542182
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K991140
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7210812S
Device Catalogue Number7210812S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2021
Initial Date Received by Manufacturer 10/20/2020
Supplement Date Received by Manufacturer10/20/2020
02/10/2021
Initial Report FDA Received Date11/12/2020
Supplement Report FDA Received Date12/11/2020
02/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-