SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
| Model Number |
72200568S |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the light source could not turn on.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h2, h3, h6.The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was a relationship found between the returned device and the reported incident.Complaint of power up failure was confirmed.Cause of power failure is a defective power supply/ballast.Unit powered up and passed functional testing with a known good power supply installed.The complaint investigation has concluded the cause of the failure to be a defective electronic component on the power supply.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
| |
|