|
The device used in treatmeant.One 14f 13 cm long abiomed introducer sheath was returned from the customer with the dilator.There were no other accessories.Visible blood was found on the sheath and inside the sideport tubing.Upon initial inspection, the hemostatic valve was observed to be damaged.The valve was inspected in more detail under 10x magnification.The damage was found to be a tear that stemmed off of the valve split line.The introducer was manually tested by introducing water through sideport of the sheath using a syringe.The introducer leaked through the valve.The tear in the hemostatic valve is the cause for the leak through the valve.The probable root cause of the tear could be off center dilator/device insertion during use.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspection.No manufacturing defects were found.Per procedure adelante s2s introducer sheath in-process and final inspection: vacuum leak test ,-perform pressure and vacuum leak test per procedure.Per procedure introducer set, silicone seal slitting once the cutting cycle is complete, the part will be placed on the machine tray, remove the seal from the tray and perform a visual inspection using a 10x microscope.At the start of each work order the first 5 acceptable seals must be provided to qa for acceptance.Split seals: if the work order is for split seals prior to providing qa with the 5 samples, visually inspect and pull test the first 3 seals per section seal pull test.If the pull tests are within specification, continue to run the first 5 samples for qa acceptance, visually inspect and submit to qa.Production may continue and seals can be "run at risk" while qa inspection is in progress.Per ifu precuations: when assembling the sheath and dilator, care must be taken to insert the dilator tip straight through the center of the valve membrane in order to prevent inadvertent puncturing of the membrane.Directions for use: slowly retract the guidewire and dilator, leaving the sheath in position.The hemostasis valve will reduce the loss of blood and the inadvertent aspiration of air through the sheath.Introduce the catheter through the hemostasis valve/sheath and advance it into position.No further follow-up is required.Based on the investigation, a capa is not required as findings did not identify a design, labeling or manufacturing non-conformity.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|