An opened phaco tip was returned for evaluation for the report of phaco tip broke off in the patient¿s eye during sculpting.A review of the device history records traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.Two photos were reviewed by the investigation site.Photo 1 shows a phaco tip broken in two pieces.Photo 2 shows a pak label with same product and lot number as reported.The phaco tip was visually inspected and deemed nonconforming, the phaco tip was broken at the ab hole in two pieces, the break was jagged, and the phaco had even wall thickness.Wear on threads, nut and flange.The complaint evaluation confirms the phaco tip was broken.The root cause for the broken phaco tip cannot be determined from this evaluation.The phaco tip visual inspection does not show any manufacturing issue that would cause the broken phaco tip.No specific action with regard to this complaint was taken because the root cause for the complaint issue cannot be determined from this evaluation.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance, such as the broken phaco tip exhibited on the returned opened sample, is removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|