| Model Number |
03P6832 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available patient information was included.Additional patient details are not available.A review of tickets determined that there is normal complaint activity for lot 72837un20.Trending review determined no adverse or non-statistical trend for falsely elevated results for the product.Returns were not available.The patient sample was retested and generated lower normal results.Qc results were reported to be within range.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no product deficiency was identified for the magnesium reagent lot 72899un20.
|
| |
|
Event or Problem Description
|
|
The customer observed a falsely elevated architect magnesium result of 3.61 mmol/l for a patient sample (sid (b)(6)) that retested at 0.73 and 0.74 mmol/l.The elevated result was not report out of the lab.No adverse impact to patient management was reported.
|
| |
|