Model Number 3854
Medical Device Problem Codes Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 10/22/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the balloon burst and blade detached.A percutaneous coronary intervention was being performed.The 4.0x12 target lesion was located in the moderately tortuous and severely calcified complex proximal right coronary artery.A 10mmx4.00mm wolverine coronary cutting balloon was selected for use.During procedure, the balloon got stuck and burst upon second inflation at 12atm for 10 seconds.The device was tried to pulled out, re-inflated and ballooned on buddy wire.The device was then trapped in a guideliner and snared out.It was also noted that the blade came in contact with plaque just proximal to stent and detached.The procedure was completed with a different device.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Common Device Name
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10830628
Report Number2134265-2020-15687
Device Sequence Number19559232
Product Code NWX
UDI-Device Identifier08714729888307
UDI-Public08714729888307
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2022
Device Model Number3854
Device Catalogue Number3854
Device Lot Number0025546644
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/22/2020
Initial Report FDA Received Date11/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-