• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD-LEGACY DUODOPA AMBULATORY INFUSION PUMP; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD-LEGACY DUODOPA AMBULATORY INFUSION PUMP; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 1400
Device Problem Use of Device Problem (1670)
Patient Problem Fatigue (1849)
Event Date 10/28/2020
Event Type  malfunction  
Event Description
Information was received indicating that a patient's wife forgot to turn off the patient's smiths medical cadd-legacy duodopa ambulatory infusion pump "at 10 p.M.Yesterday because she took strong painkillers and fell asleep." it was reported that medication was delivered for twenty-four (24) hours instead of sixteen (16) hours; and that approximately 36ml was over-delivered.Per reporter, the programmed settings on the pump were: morning dose 12, continuous day rate 4.5 and extra dose 0.5.It was reported that the patient was "a little tired" and no additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-LEGACY DUODOPA AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key10831031
MDR Text Key216083734
Report Number3012307300-2020-11403
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1400
Device Catalogue Number21-1400-03
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age73 YR
-
-