MEDOS INTERNATIONAL SARL HD EPSCP,4.0,30,167,MITEK; RIGID ENDOSCOPE Back to Search Results
Model Number 242018
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/27/2020
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that the endoscope device appeared to have spots on its lens.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the device was received and evaluated at the service center.The reported complaint that the scope had spots on the lenses, was confirmed.The following defects were found with the device upon evaluation : - outer tube damaged, distal tip damaged distal tip has deposits.- image error, on camera-image cloudy/blurred.The defective components were replaced and the device was tested and found to be fully functional.The damage to the distal tip is a result of contact with the shaver during a surgical process, as identified during evaluation.The damage to the outer tube can be attributed to user mishandling of the device or a probable fall.Improper cleaning / maintenance from the customer would have caused the deposits on the distal tip.The defective distal tip would have caused the poor image display by the scope.A manufacturing record evaluation was performed for the finished device [serial : 1376753], and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 correction narrative: d9: the date device returned to manufacturer has been updated to reflect the correct information.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD EPSCP,4.0,30,167,MITEK
Type of Device
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
MDR Report Key10831250
MDR Text Key215986087
Report Number1221934-2020-03457
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705029150
UDI-Public10886705029150
Combination Product (y/n)N
PMA/PMN Number
K080560
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number242018
Device Catalogue Number242018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/12/2020
Supplement Dates Manufacturer Received12/03/2020
12/24/2020
Supplement Dates FDA Received12/04/2020
12/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-