• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 SCREW HOLE COVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 SCREW HOLE COVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71369894
Device Problems Inadequacy of Device Shape and/or Size (1583); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/20/2020
Event Type  malfunction  
Event Description
It was reported that during surgery, the screw did not fit into the cup.The procedure was completed using a s+n backup device.A surgical delay of 10 minutes was reported.No injury to the patient was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage from attempted use.The device was manufactured in 2017.A dimensional evaluation could not confirm the stated failure mode.The device was found to be within specification.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.The potential probable cause for this event is likely improper insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 SCREW HOLE COVER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10832533
MDR Text Key216014860
Report Number1020279-2020-06406
Device Sequence Number1
Product Code MBL
UDI-Device Identifier03596010607010
UDI-Public03596010607010
Combination Product (y/n)N
PMA/PMN Number
K070756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71369894
Device Catalogue Number71369894
Device Lot Number17LM10735J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Initial Date Manufacturer Received 10/21/2020
Initial Date FDA Received11/12/2020
Supplement Dates Manufacturer Received04/16/2021
Supplement Dates FDA Received05/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-