SMITH & NEPHEW, INC. FRAME STABILIZER TOOL; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
|
Back to Search Results |
|
| Model Number |
71070343 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the 10mm nut is cross threaded on the frame stabilization clamp.No case or patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the threads are damaged on the nut of the frame stabilizer tool.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|
|
|