|
The device, used in treatment, was returned for evaluation.A visual inspection of the returned cutting block confirms the device broke into several pieces.The cutting block, spring, spiked cross bar, cross bar screw and cutting block knob were returned.The posterior ref indicator pin and cup point set screw was not returned with the other pieces.This device was manufactured in 2016.This device exhibits signs of significant wear/usage.A medical investigation was conducted and this case reports that the adjustment knob and back came off of the cutting block during use.A photo of the broken device confirms not all pieces were recovered.Per complaint detail, there was no patient injury reported.It is unknown whether a backup device was required, or if there was a surgical delay.Since there is no patient harm alleged, no further clinical assessment is warranted.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
|