SMITH & NEPHEW, INC. MOT DR UNIT HNDCONT POWERMAX ELITE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Model Number 72200872 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Event Description
|
It was reported that the mdu had an error message when plugged into box.When testing shaver prior to case hand unit got warm.No case involved.Therefore, there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The gearbox was found to be jammed.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A blade stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|