• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSURE SENSE LEAD MRI SURESCAN; ELECTRODE, PACEMAKER, PERMANENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSURE SENSE LEAD MRI SURESCAN; ELECTRODE, PACEMAKER, PERMANENT Back to Search Results
Model Number 407452
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/10/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that infection occurred in the implantable pulse generator (ipg) pocket.Antibiotics were administered.The system was explanted and was replaced with a leadless implantable pulse generator (ipg).No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SENSE LEAD MRI SURESCAN
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key10834638
Report Number2649622-2020-22121
Device Sequence Number17025542
Product Code DTB
UDI-Device Identifier00643169590519
UDI-Public00643169590519
Combination Product (Y/N)Y
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
P830061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/02/2018
Device Model Number407452
Device Catalogue Number407452
Was Device Available for Evaluation? No
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 11/12/2020
Initial Report FDA Received Date11/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
A3SR01 IPG
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
-
-