Model Number 7210707
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Information (3190)
Date of Event 10/21/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the surgery, the technique was being carried out step by step, at the time the suturing started at the second point the twinfix titanium threads got disengaged from the needle, this happened with the white and blue sutures at the 4 ends.The suturing could not be performed.No significant delay or other complications were reported.Smith and nephew twinfix back up device was used to complete the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Event or Problem Description
It was reported that during a ans/ small joints repair procedure, the technique was being carried out step by step, at the time the suturing started at the second point the twinfix titanium threads got disengaged from the needle, this happened with the white and blue sutures at the 4 ends.The suturing could not be performed.No significant delay or other complications were reported.Smith and nephew twinfix back up device was used into an additional bone hole to complete the surgery.Anchors were left in the patient.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWNFX TI 3.5 STR ANCHR W/2 ULTRB STR N
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10835542
Report Number1219602-2020-01836
Device Sequence Number17019518
Product Code MBI
UDI-Device Identifier03596010532190
UDI-Public03596010532190
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K972326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/03/2025
Device Model Number7210707
Device Catalogue Number7210707
Device Lot Number2053678
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/22/2020
Supplement Date Received by Manufacturer04/19/2021
Initial Report FDA Received Date11/13/2020
Supplement Report FDA Received Date04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age20 YR
Patient Weight85
-
-