| Model Number |
7205357 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the mini motor drive unit stalled.It is unknown whether the event happened during surgery and if there was a patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that the mini motor drive unit stalled.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).Event: updated.The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review concluded this was a repeat issue.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and a connector pin was pushed back into the connector.Complaint of motor stall could not be reproduced.Functional testing could not be performed due to damaged connector pin.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
| |
|