MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Obstruction of Flow (2423); Infusion or Flow Problem (2964)
Patient Problems Muscular Rigidity (1968); Therapeutic Effects, Unexpected (2099)
Event Date 09/18/2020
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2020, product type: catheter.Product id: 8780, serial/lot #: (b)(4), ubd: 21-oct-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) via a company representative regarding a patient receiving lioresal (1000 mcg/ml at 138 mcg/day) via an implanted pump.It was reported that the patient had not had good efficacy post implant.The dose had been increased from 80 mcg/day to 138 mcg/day with no relief from spasticity.A dye study showed there was no dye at the end of the catheter.The catheter was reported to be ¿bad¿ and ¿likely occluded¿.The physician wanted to do a rotor study and go from there.It was unknown if there were any environmental, external, or patient factors that may have led or contributed to the issue.The issue was not resolved, and it was indicated that the hcp had no further information to provide regarding the event.The patient was scheduled for surgery on (b)(6) 2020.
 
Manufacturer Narrative
H3: analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.The returned pump was interrogated and as per the logs gablofen with concentration 1000 mcg/ml was being administered at a dose rate of 79.8 mcg/day as of (b)(6) 2020.This is discrepant with previous information having reported that the pump administered lioresal.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received via a company representative.The pump and catheter were returned to the manufacturer.
 
Manufacturer Narrative
Continuation of d10: product id 8780 lot# serial# (b)(6), implanted: (b)(6) 2020, explanted: (b)(6) 2020, product type catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a healthcare professional (hcp) via a company representative who reported that the catheter occlusion was not confirmed during the procedure on (b)(6) 2020.The catheter drawback of csf (cerebrospinal fluid) was fine and the pump was running fine.The cause of the inconclusive dye study was unknown.The pump and catheter were replaced during the procedure because the hcp didn¿t want to take a chance on anything.
 
Manufacturer Narrative
H3: analysis of the implantable intrathecal catheter (b)(6) found no significant anomalies.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10835594
MDR Text Key216110384
Report Number3004209178-2020-19925
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169630505
UDI-Public00643169630505
Combination Product (y/n)N
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2021
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Initial Date Manufacturer Received 11/10/2020
Initial Date FDA Received11/13/2020
Supplement Dates Manufacturer Received11/13/2020
12/11/2020
02/05/2021
Supplement Dates FDA Received11/18/2020
12/15/2020
02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age22 YR
-
-