MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Obstruction of Flow (2423); Infusion or Flow Problem (2964)
|
Patient Problems
Muscular Rigidity (1968); Therapeutic Effects, Unexpected (2099)
|
Event Date 09/18/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2020, product type: catheter.Product id: 8780, serial/lot #: (b)(4), ubd: 21-oct-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare professional (hcp) via a company representative regarding a patient receiving lioresal (1000 mcg/ml at 138 mcg/day) via an implanted pump.It was reported that the patient had not had good efficacy post implant.The dose had been increased from 80 mcg/day to 138 mcg/day with no relief from spasticity.A dye study showed there was no dye at the end of the catheter.The catheter was reported to be ¿bad¿ and ¿likely occluded¿.The physician wanted to do a rotor study and go from there.It was unknown if there were any environmental, external, or patient factors that may have led or contributed to the issue.The issue was not resolved, and it was indicated that the hcp had no further information to provide regarding the event.The patient was scheduled for surgery on (b)(6) 2020.
|
|
Manufacturer Narrative
|
H3: analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.The returned pump was interrogated and as per the logs gablofen with concentration 1000 mcg/ml was being administered at a dose rate of 79.8 mcg/day as of (b)(6) 2020.This is discrepant with previous information having reported that the pump administered lioresal.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received via a company representative.The pump and catheter were returned to the manufacturer.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 8780 lot# serial# (b)(6), implanted: (b)(6) 2020, explanted: (b)(6) 2020, product type catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a healthcare professional (hcp) via a company representative who reported that the catheter occlusion was not confirmed during the procedure on (b)(6) 2020.The catheter drawback of csf (cerebrospinal fluid) was fine and the pump was running fine.The cause of the inconclusive dye study was unknown.The pump and catheter were replaced during the procedure because the hcp didn¿t want to take a chance on anything.
|
|
Manufacturer Narrative
|
H3: analysis of the implantable intrathecal catheter (b)(6) found no significant anomalies.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|